ABB TTH200 Head Mounted Temperature Transmitter
₹15,600.00
(GST 18% Exclusive)
1792
(However please contact store for precise delivery schedule. Backorder also allowed)
ABB TTH200 Head Mounted Temperature Transmitter is for standard B sensor heads. This High-Range Temperature Transmitter is provided for all HART standard applications.
Applications: Temperature measurement in air, soil, or water, measurement of delicate surfaces (e.g., leaves, fruits), Oil, Gas and Petrochemical, Research and Development, Hazardous Areas.
Customers who bought this item also bought
-
₹3,376.00 – ₹6,912.00
(GST 18% Exclusive)
Novus Din Rail Temperature Transmitter
-
₹14,900.00
(GST 18% Exclusive)
Siemens QFA3160 Humidity and Temperature Sensor
-
₹2,018.00
(GST 18% Exclusive)
Ace AI-RTD-FLP Flameproof Temperature Sensor
-
₹5,200.00
(GST 18% Exclusive)
Novus USB Head Mounted Temperature Transmitter
-
₹15,038.14
(GST 18% Exclusive)
ABB TTH300 Head Mounted Temperature Transmitter
-
₹1,397.00
(GST 18% Exclusive)
Ace AI-TTx Temperature Transmitter
ABB TTH200 Head Mounted Temperature Transmitter is designed for standard type B sensor heads. The ABB TTH200 is provided for all HART standard applications in temperature measurement with a single sensor. This Head Mount Temperature Sensor can support the functionality as same as ABB TTH 300 Temperature Transmitter but the difference is that without sensor drift and sensor redundancy monitoring that can be configured via any standard FTD tool via EDD or FDT / DTM in DAT200 (Asset Vision Basic).
Features Of Head-Mounted Temperature Transmitter:
- HART Compatibility
- ATEX, IECEx, Zone 0 Compliance
- The TTH200 meets all requirements for the CE mark by IEC 61326 (2006)
- Conforms with IEC 61508 as regards use in safety-related applications, up to and including SIL 2.
- Event monitor for logging critical events
- Universal sensor input for resistance thermometers (RTD) and thermocouples
Additional information
Manufacturer | ABB |
---|---|
Model No | TTH 200 |
Output | Standard: 4 to 20 mA; Configurable: 20 to 4 mA |
Cable Connection | Maximum sensor line resistance (RW) for each line: 1.5 kΩ, total: 3 kΩ |
Storage Temperature | -50 to 85 °C |
Electrical Connection | Electrical connection Terminals (captive screws) incl. soldering tags |
Input | Resistance RTD Pt-100, Thermocouples B,E, J, K, N, R,S, T, Voltages, mV Voltages |
Voltage | -125 to 125 mV; -125 to 1100 mV |
Ambient Temperature | Standard: -40 to 85°C; Optional: -50 to 85°C |
Electrical Interface | 3.5 kV DC (approx. 2.5 kV AC) 60 s, input to output |
Vibration Resistance | 10 to 2,000 Hz at 5g in acc. with IEC 60068-2-6, during operation and transport |
Input resistance | > 10 MΩ |
Display | Standard Supply Without Display |
Warm-up Time | 5 minutes |
Power Supply | 11 to 42 VDC |
Housing | Polycarbonate |
Communication | HART protocol |
Calibration | Provided alongwith and valid for 1 year, traceable to National Standards. |
Weight | 50 Grams |