Hioki DT4261 Digital Multimeter

Product Id : 281254

17,175.00

Hioki DT4261 Digital Multimeter can be connected to a computer via USB, allowing measurement data to be transferred to Hioki’s software for further analysis and reporting. Overall, the Hioki 4261 is a high-precision multimeter that is suitable for a wide range of electrical measurement applications. Its accuracy, safety features, and data logging capabilities make it a useful tool for professionals and hobbyists alike.

Applications: Electrical maintenance and troubleshooting, HVAC system maintenance, Automotive repair, Laboratory research, Electronics manufacturing, Renewable energy systems.



  HTTPS Secured PaymentHTTPS, the lock icon in the address bar, an encrypted website connection. While it was once reserved primarily for passwords and other sensitive data, the entire web is gradually leaving HTTP behind and switching to HTTPS. The 'S' in HTTPS stands for 'Secure'. It’s the secure version of the standard “hypertext transfer protocol”. Most importantly, your payment is also secured.

   Bulk Purchase? Reach us for special price
  Free shipping all over India
  Ask a question?

Country : United States

Product Description

Hioki DT4261 digital multimeter is designed to measure various electrical parameters accurately and efficiently. It has a large, backlit display that makes it easy to read measurements even in low light conditions. IN addition to its accuracy, the Hioki DT4261 digital Multimeter also has various safety features. It is designed to meet safety standards such as CAT IV 600V, CAT III 1000V, and EN61010-1. It also has an auto power-off function to conserve battery life. The meter also has a data logging function that can record up to 1000 sets of measurement data. This makes it useful for applications where it is necessary to track and analyze multiple measurements over time.

Features of Hioki DT4261 Digital Multimeter:

  • Measurement functions: The Hioki 4261 can measure AC/DC voltage, AC/DC current, resistance, capacitance, frequency, temperature, and continuity.
  • High accuracy: The meter has a basic accuracy of 0.05% and a maximum resolution of 1μV, 0.01μA, 0.01Ω, and 0.001nF.
  • Safety features: The Hioki 4261 is designed to meet safety standards such as CAT IV 600V, CAT III 1000V, and EN61010-1. It also has an auto power-off function to conserve battery life.
  • Data logging: The meter has a data logging function that can record up to 1000 sets of measurement data.
  • PC connectivity: The Hioki 4261 can be connected to a computer via USB to transfer measurement data to Hikoi’s software for analysis and reporting.
  • Backlit display: The meter has a backlit display for easy reading in low light conditions.

 

Additional Information

Manufacturer Hioki
Model No DT4261
Measuring Parameters AC/DC voltage, AC/DC current, resistance, capacitance, frequency, temperature, and continuity.
DC Voltage 600.0 mV to 1000 V, 5 ranges, Basic accuracy: ±0.15% rdg. ±2 dgt.
AC Voltage 6.000 V to 1000 V, 4 ranges, Frequency characteristics: 40 Hz to 1 kHz Basic accuracy 40 Hz - 500 Hz: ±0.9% rdg. ±3 dgt. (True RMS, crest factor 3 or less)
DC + AC Voltage range 6.000 V to 1000 V, 4 ranges, Frequency characteristics: DC, 40 Hz to 1 kHz Basic accuracy DC, 40 Hz - 500 Hz: ±1.0% rdg. ±13 dgt. (True RMS, crest factor 3 or less)
LoZ V 600.0 V, 1 range, Frequency characteristics: DC, 40 Hz to 1 kHz Basic accuracy DC, 40 Hz - 500 Hz: ±1.0% rdg. ±13 dgt. (True RMS, crest factor 3 or less)
Resistance 600.0 Ω to 60.00 MΩ, 6 ranges, Basic accuracy: ±0.7% rdg. ±3 dgt.
DC Current 600.0 mA to 10.00 A, 3 ranges Basic accuracy: ±0.5% rdg. ±3 dgt.
AC Current 600.0 mA to 10.00 A, 3 ranges Basic accuracy 40 Hz - 500 Hz: ±1.4% rdg. ±3 dgt. (True RMS, crest factor 3 or less) Frequency characteristics: 40 Hz to 1 kHz
AC Current range (use with Clamp on probes) 10.00 A to 1000 A, 7 ranges Basic accuracy 40 Hz - 500 Hz: Add the Clamp on probe accuracy to ±0.9% rdg. ±3 dgt. (True RMS, crest factor 3 or less)
Capacitance 1.000 μF to 10.00 mF, 5 ranges, Basic accuracy: ±1.9% rdg. ±5 dgt.
Continuity Check Continuity threshold ON : 25 Ω, Continuity threshold OFF : 245 Ω, Response time: 0.5 ms or more
Diode Test Open terminal voltage: 2.0 V or less, Testing current: 0.2 mA or less, Threshold of forward voltage: 0.15 V to 1.8 V
Voltage Frequency Range 99.99 Hz to 99.99 kHz, 4 ranges (Limited by minimum sensitivity voltage) Basic accuracy: ±0.1% rdg. ±1 dgt.
Current frequency range 99.99 Hz to 9.999 kHz, 3 ranges (Limited by minimum sensitivity current) Basic accuracy: ±0.1% rdg. ±1 dgt.
Display Main and sub displays: 4-digits LCD, max. 6000 digits (excluding frequency measurement), bar-graph
Display refresh rates 5 times/s (Capacitance measurement: 0.05 to 5 times/s, depending on measured value, Frequency: 1 to 2 times/s)
Power Supply LR6 (AA) alkaline batteries × 3, Continuous operating time: 130 hr. (without Z3210 installed), 70 hr. (with Z3210 installed and using wireless communications)
Supply Scope Test Lead L9300 × 1, Instruction Manual × 1, LR6 (AA) alkaline battery × 3, Operating Precautions ×1
Dimension 87 mm (3.43 in.) W × 185 mm (7.28 in.) H × 47 mm (1.85 in.) D, 480 g (16.9 oz.) (with test leads holder and batteries)